As a Authorized dealer we are engaged in providing the broad assortment of wide gamut of optical instruments for various applications. Densitometer measure the amount of transmitted light for radio graphic testing developed and known for accurate reading, favorable stability and convenient operation are its user friendly features. It can be widely used in various industries, including nondestructive testing (NDT), Printing and Medical treatment.
3000 series black-white densitometer is used for measuring the blackness of films; it has advantages of accurate reading, high stability, strong anti-interference capacity, attractive appearance and easy operation. It can be widely used in the fields such as nondestructive testing, sensitive materials, printing industry and medical industry etc. each unit is equipped with density tablet and calibration software for testing its accuracy.
The inclusive density film with the machine is used for calibration.
Specification
Density Range | D=0.00 -4.00 (DM3010) |
Repeatability | ±0.02 |
Measuring Error | 0.00 -2.00 0.02 |
Power | 50 ~ 60Hz, 100V-120V or 200V-240V |
Power Dissipation | 25W |
Aperture Size | Dia. 2mm |
Display | Three digital display |
Light Source | 27pcs LED |
Sampling Time | 0.8 s |
Communication Port | RS232 Port, optional USB connection |
Working Environment | Temperature 0 ~40 , Relative humidity(RH) 85% |
Dimension | 260 x 265 x 110 (H x W x L) |
Weight | 2.1kg |
Includes:
Unique calibration software which allows precise user calibration to a certified step wedge.
How to Calibration
Note: The calibration is effectively only after restart the densitometer when 3 appears.
The meaning of densitometer window shows
E.1 | Zero Prompt |
E.2 | Exceed the measurements limit prompt |
E.3 | Calibration download success prompt |
E.4 | Delete calibration data success prompt |
E.5 | Read the device version number success prompt |
Black-white densitometer is a precision measuring instrument. It should be keep away from dusts, humidity, alkaline and acid objects, and magnetic fields.
The dust is easy to fall in to the light aperture, so after a period of time, if the instrument donot display 0.00 when calibrating, please open the light source shell with tape, then gently wipe the dust on the light-emitting diode with a cotton swab dipped in alcohol to ensure the normal use of the instrument.
When the density values are out of the prescribed range, the measurement data are still has reference value.
Scope Of Supply :
Film density is measured with a densitometer which simply measures the amount of light transmitted through a piece of film using photovoltic sensor